Vitros immunodiagnostic products hiv combo reagent pack

Official notice

14 Sept 2022

Last checked

14 Sept 2022

Medium riskHealth Canada

An official Health Canada notice concerns Vitros immunodiagnostic products hiv combo reagent pack. Scenario 1: Ortho received a customer complaint for false negative results generated from Vitros HIV combo reagent pack, lot 0670, when testing a proficiency sample identified to be reactive for the presence of HIV-1 p24 antigen and absent of antibodies to HIV (p24 antigen positive). Ortho's investigation confirmed this issue and identified five additional lots of Vitros HIV combo reagent pack that also generated false negative results when tested using the same hiv-1 p24 positive sample. For this scenario, to date, Ortho's investigation has shown that this issue is limited only to Hiv-1 p24 antigen detection and does not affect the antibody detection of antihiv-1 or anti-hiv-2 in Vitros Hiv combo reagent pack. Scenario 2: During the investigation in scenario 1, Ortho received complaints related to repeat reactive results for Vitros Hiv combo reagent pack. Ortho confirmed that known negative samples could produce reactive results. Ortho's data demonstrates a correlation to the same lots implicated in scenario 1. Recall start date: August 30, 2022 Published 2022-09-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64548

Official notice

2022-09-14

The hazard

Scenario 1: Ortho received a customer complaint for false negative results generated from Vitros HIV combo reagent pack, lot 0670, when testing a proficiency sample identified to be reactive for the presence of HIV-1 p24 antigen and absent of antibodies to HIV (p24 antigen positive). Ortho's investigation confirmed this issue and identified five additional lots of Vitros HIV combo reagent pack that also generated false negative results when tested using the same hiv-1 p24 positive sample. For this scenario, to date, Ortho's investigation has shown that this issue is limited only to Hiv-1 p24 antigen detection and does not affect the antibody detection of antihiv-1 or anti-hiv-2 in Vitros Hiv combo reagent pack. Scenario 2: During the investigation in scenario 1, Ortho received complaints related to repeat reactive results for Vitros Hiv combo reagent pack. Ortho confirmed that known negative samples could produce reactive results. Ortho's data demonstrates a correlation to the same lots implicated in scenario 1. Recall start date: August 30, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

0670, when

Recall Numbers

64548

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Vitros Immunodiagnostic Products HIV Combo Reagent Pack, Lot or serial number - 0670 0740 0760 0660 0730 0750, Model or catalog number - 684 2779

Risk / hazard

Scenario 1: Ortho received a customer complaint for false negative results generated from Vitros HIV combo reagent pack, lot 0670, when testing a proficiency sample identified to be reactive for the presence of HIV-1 p24 antigen and absent of antibodies to HIV (p24 antigen positive). Ortho's investigation confirmed this issue and identified five additional lots of Vitros HIV combo reagent pack that also generated false negative results when tested using the same hiv-1 p24 positive sample. For this scenario, to date, Ortho's investigation has shown that this issue is limited only to Hiv-1 p24 antigen detection and does not affect the antibody detection of antihiv-1 or anti-hiv-2 in Vitros Hiv combo reagent pack. Scenario 2: During the investigation in scenario 1, Ortho received complaints related to repeat reactive results for Vitros Hiv combo reagent pack. Ortho confirmed that known negative samples could produce reactive results. Ortho's data demonstrates a correlation to the same lots implicated in scenario 1. Recall start date: August 30, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Ortho-clinical Diagnostics, Vitros Immunodiagnostic Hiv Combo Reagent Pack.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

64548

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 Sept 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.