No image published for this recall

No image published for this recall

Vitros® eci immunodiagnostic system - troponin i es reagent pack

Official notice

22 Jun 2023

Last checked

22 Jun 2023

Medium riskHealth Canada

An official Health Canada notice concerns Vitros® eci immunodiagnostic system - troponin i es reagent pack. Ortho Clinical Diagnostics (QuidelOrtho) became aware of an issue affecting VITROS® Immunodiagnostic products Troponin I Es Reagent Pack, lot 4800. Due to a manufacturing error, around 1% of reagent packs contain incorrect wells which will generate inaccurate results. During a review of manufacturing records, QuidelOrtho confirmed that 87 incorrect wells (0.022% of total wells produced) were included in the manufacture of VITROS® Troponin I Es Reagent Pack, lot 4800. If the affected wells are used for calibration, quality control testing, or sample testing, they will not produce the appropriate or expected signal levels. Please note that while a reagent pack may include multiple affected wells, not all wells within the reagent pack are affected. Recall start date : June 7, 2023 Published 2023-06-22.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73864

Official notice

2023-06-22

The hazard

Ortho Clinical Diagnostics (QuidelOrtho) became aware of an issue affecting VITROS® Immunodiagnostic products Troponin I Es Reagent Pack, lot 4800. Due to a manufacturing error, around 1% of reagent packs contain incorrect wells which will generate inaccurate results. During a review of manufacturing records, QuidelOrtho confirmed that 87 incorrect wells (0.022% of total wells produced) were included in the manufacture of VITROS® Troponin I Es Reagent Pack, lot 4800. If the affected wells are used for calibration, quality control testing, or sample testing, they will not produce the appropriate or expected signal levels. Please note that while a reagent pack may include multiple affected wells, not all wells within the reagent pack are affected. Recall start date : June 7, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73864

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - VITROS® ECi Immunodiagnostic System - Troponin I Es Reagent Pack, Lot or serial number - 4800, Model or catalog number - 680 2301 680 2302

Risk / hazard

Ortho Clinical Diagnostics (QuidelOrtho) became aware of an issue affecting VITROS® Immunodiagnostic products Troponin I Es Reagent Pack, lot 4800. Due to a manufacturing error, around 1% of reagent packs contain incorrect wells which will generate inaccurate results. During a review of manufacturing records, QuidelOrtho confirmed that 87 incorrect wells (0.022% of total wells produced) were included in the manufacture of VITROS® Troponin I Es Reagent Pack, lot 4800. If the affected wells are used for calibration, quality control testing, or sample testing, they will not produce the appropriate or expected signal levels. Please note that while a reagent pack may include multiple affected wells, not all wells within the reagent pack are affected. Recall start date : June 7, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Ortho-clinical Diagnostics, Vitros® Eci Immunodiagnostic System - Troponin I Es Reagent Pack.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

73864

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 22 Jun 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.