Ventana pd-l1 (sp142)

Official notice

14 Dec 2022

Last checked

14 Dec 2022

Critical riskHealth Canada

An official Health Canada notice concerns Ventana pd-l1 (sp142). During internal comparison studies, unacceptable light staining with some VENTANA PD-L1 (SP142) on-markets lots were confirmed which affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result. Recall start date: November 28, 2022 Published 2022-12-14.

What to do now

Roche diagnostics . The roche support network customer support center at 1-800-227-2155.

Recall number

72338

Official notice

2022-12-14

The hazard

During internal comparison studies, unacceptable light staining with some VENTANA PD-L1 (SP142) on-markets lots were confirmed which affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result. Recall start date: November 28, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72338

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - VENTANA PD-L1 (SP142), Lot or serial number - H07164 F31427 J12769 J14244 H23050 G16831 G06715 j18775 J00572 H15132 J18775 H00001 H000001, Model or catalog number - 741-4860 740-4859

Risk / hazard

During internal comparison studies, unacceptable light staining with some VENTANA PD-L1 (SP142) on-markets lots were confirmed which affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result. Recall start date: November 28, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ventana Medical, Ventana Pd-l1.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

72338

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 Dec 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.