No image published for this recall

No image published for this recall

Varisoft

Official notice

1 Nov 2023

Last checked

1 Nov 2023

Medium riskHealth Canada

An official Health Canada notice concerns Varisoft. Unomedical A/S has found that in rare cases the VariSoft infusion set connector detaches more easily from the infusion set than expected, requiring less force to disconnect than intended, thereby interrupting the delivery of insulin. Recall start date: October 11, 2023 Published 2023-11-01.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74511

Official notice

2023-11-01

The hazard

Unomedical A/S has found that in rare cases the VariSoft infusion set connector detaches more easily from the infusion set than expected, requiring less force to disconnect than intended, thereby interrupting the delivery of insulin. Recall start date: October 11, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74511

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - VariSoft, Lot or serial number - 5388368 5388372 5388366 5388371 5388376 5388357 5388362 5388367, Model or catalog number - 1002827 1002828 1001681 1002830 1002832

Risk / hazard

Unomedical A/S has found that in rare cases the VariSoft infusion set connector detaches more easily from the infusion set than expected, requiring less force to disconnect than intended, thereby interrupting the delivery of insulin. Recall start date: October 11, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Unomedical A S, Varisoft.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74511

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 1 Nov 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.