No image published for this recall

No image published for this recall

Vantage galan 3t

Official notice

12 Nov 2025

Last checked

12 Nov 2025

Medium riskHealth Canada

An official Health Canada notice concerns Vantage galan 3t. It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak. Published 2025-11-12.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78458

Official notice

2025-11-12

The hazard

It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78458

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Vantage Galan 3T, Lot or serial number - Y5C1922001, Model or catalogue number - MRT-3020/Y5

Risk / hazard

It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Canon, Vantage Galan 3t.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

78458

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 12 Nov 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.