No image published for this recall

No image published for this recall

Trusight cystic fibrosis library prep kit

Official notice

13 Jun 2022

Last checked

13 Jun 2022

Medium riskHealth Canada

An official Health Canada notice concerns Trusight cystic fibrosis library prep kit. Annotation error in analysis software - the genomic coordinate reported for one variant in the sample did not match the cDNA name; the variant corresponding to the genomic coordinate is not cystic fibrosis-causing, but the cDNA name is potentially cystic fibrosis-causing (software issue/malfunction). Affected device details: catalog #20036925, affected software: local run manager (lrm) cf clinical seq 2.0 analysis module, software version:1.0.0.56, affected report file: trusightcfclinicalsequencingassay.txt. Recall start date: May 25, 2022 Published 2022-06-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64268

Official notice

2022-06-13

The hazard

Annotation error in analysis software - the genomic coordinate reported for one variant in the sample did not match the cDNA name; the variant corresponding to the genomic coordinate is not cystic fibrosis-causing, but the cDNA name is potentially cystic fibrosis-causing (software issue/malfunction). Affected device details: catalog #20036925, affected software: local run manager (lrm) cf clinical seq 2.0 analysis module, software version:1.0.0.56, affected report file: trusightcfclinicalsequencingassay.txt. Recall start date: May 25, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64268

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Trusight Cystic Fibrosis Library Prep Kit, Lot or serial number - 00816270020149, Model or catalog number - 20036925

Risk / hazard

Annotation error in analysis software - the genomic coordinate reported for one variant in the sample did not match the cDNA name; the variant corresponding to the genomic coordinate is not cystic fibrosis-causing, but the cDNA name is potentially cystic fibrosis-causing (software issue/malfunction). Affected device details: catalog #20036925, affected software: local run manager (lrm) cf clinical seq 2.0 analysis module, software version:1.0.0.56, affected report file: trusightcfclinicalsequencingassay.txt. Recall start date: May 25, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ilumina, Trusight Cystic Fibrosis Library Prep Kit.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

64268

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Airseal access port with blunt obturator and airseal access port with optical obturator

Medium riskHealth Canada2 Jul 2026
CA
No image published for this recall

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
No image published for this recall

M zn-air battery pak

Medium riskHealth Canada6 May 2026
CA
No image published for this recall

Exent® analyser

Medium riskHealth Canada29 Apr 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Jun 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.