No image published for this recall
Trilogy evo and ev300
Official notice
22 Mar 2024
Last checked
22 Mar 2024
An official Health Canada notice concerns Trilogy evo and ev300. Philips Respironics has become aware of a potential safety issue with Trilogy EVO and Trilogy EV300 devices where the ventilator can issue a battery depleted or loss of power alarm while sufficient power is still available. This can result in a sudden loss of ventilation while the device alarms. A software algorithm that calculates remaining battery life can malfunction and cause the device to either: 1. Issue a loss of power alarm that stops CPAP or PSV therapy while operating on battery power alone. 2. Issue a battery depleted alarm while continuing therapy if plugged into a permanent power source. Recall start date: March 4, 2024 Published 2024-03-22.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75281
Official notice
2024-03-22
The hazard
The hazard
Philips Respironics has become aware of a potential safety issue with Trilogy EVO and Trilogy EV300 devices where the ventilator can issue a battery depleted or loss of power alarm while sufficient power is still available. This can result in a sudden loss of ventilation while the device alarms. A software algorithm that calculates remaining battery life can malfunction and cause the device to either: 1. Issue a loss of power alarm that stops CPAP or PSV therapy while operating on battery power alone. 2. Issue a battery depleted alarm while continuing therapy if plugged into a permanent power source. Recall start date: March 4, 2024
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips, Trilogy Evo And Ev300.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 22 Mar 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.