Trexo home and trexo plus

Official notice

24 Aug 2023

Last checked

24 Aug 2023

Medium riskHealth Canada

An official Health Canada notice concerns Trexo home and trexo plus. The following malfunctions and nonconformances were observed during Trexo lab-testing and Trexo design reviews of the device, Resulting in this voluntary device recall and proposed field correction. Electromagnetic (EM) radiation levels above threshold (Ref: IEC 60601-1-11), as observed during lab-testing contracted by Trexo. Water ingress was observed within electronics enclosures after IP22 testing, as observed during lab-testing contracted by Trexo. Recall start date: July 31, 2023 Published 2023-08-24.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74192

Official notice

2023-08-24

The hazard

The following malfunctions and nonconformances were observed during Trexo lab-testing and Trexo design reviews of the device, Resulting in this voluntary device recall and proposed field correction. Electromagnetic (EM) radiation levels above threshold (Ref: IEC 60601-1-11), as observed during lab-testing contracted by Trexo. Water ingress was observed within electronics enclosures after IP22 testing, as observed during lab-testing contracted by Trexo. Recall start date: July 31, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74192

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Trexo Home, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - A0001

Product 2

Affected products - Trexo Plus, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - A0001

Risk / hazard

The following malfunctions and nonconformances were observed during Trexo lab-testing and Trexo design reviews of the device, Resulting in this voluntary device recall and proposed field correction. Electromagnetic (EM) radiation levels above threshold (Ref: IEC 60601-1-11), as observed during lab-testing contracted by Trexo. Water ingress was observed within electronics enclosures after IP22 testing, as observed during lab-testing contracted by Trexo. Recall start date: July 31, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Trexo Robotics, Trexo Home And Trexo Plus.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

74192

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 24 Aug 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.