No image published for this recall

No image published for this recall

Titan® inflatable penile prosthesis

Official notice

10 Nov 2023

Last checked

10 Nov 2023

Medium riskHealth Canada

An official Health Canada notice concerns Titan® inflatable penile prosthesis. Pump bulbs are manufactured using 2 molds. A component on one of the two molds shifted during production. This resulted in a gradual build of material on the mold, which gradually decreased wall thickness of the pump. This occurred over the course of several weeks, culminating in a period of time (17 September 2022 to 02 December 2022) where pump bulbs were manufactured below the wall thickness specification. Recall Start Date: October 30, 2023 Published 2023-11-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74577

Official notice

2023-11-10

The hazard

Pump bulbs are manufactured using 2 molds. A component on one of the two molds shifted during production. This resulted in a gradual build of material on the mold, which gradually decreased wall thickness of the pump. This occurred over the course of several weeks, culminating in a period of time (17 September 2022 to 02 December 2022) where pump bulbs were manufactured below the wall thickness specification. Recall Start Date: October 30, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74577

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Titan® Inflatable Penile Prosthesis, Lot or serial number - 8806754, Model or catalog number - ES2920

Risk / hazard

Pump bulbs are manufactured using 2 molds. A component on one of the two molds shifted during production. This resulted in a gradual build of material on the mold, which gradually decreased wall thickness of the pump. This occurred over the course of several weeks, culminating in a period of time (17 September 2022 to 02 December 2022) where pump bulbs were manufactured below the wall thickness specification. Recall Start Date: October 30, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Coloplast A S, Titan® Inflatable Penile Prosthesis.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

74577

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 10 Nov 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.