No image published for this recall

No image published for this recall

T:slim x insulin pumps

Official notice

8 Aug 2025

Last checked

8 Aug 2025

High riskHealth Canada

An official Health Canada notice concerns T:slim x insulin pumps. Tandem Diabetes Care, Inc. ("Tandem") has identified certain speaker versions that are in t slim X insulin pumps have a higher rate of malfunction 16 events due to a wiring issue within the speaker. Recall start date: August 18, 2025 Published 2025-08-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77883

Official notice

2025-08-08

The hazard

Tandem Diabetes Care, Inc. ("Tandem") has identified certain speaker versions that are in t slim X insulin pumps have a higher rate of malfunction 16 events due to a wiring issue within the speaker. Recall start date: August 18, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77883

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - t:slim X2 / Basal-IQ (Mmol/L), Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 1006535

Product 2

Affected products - t:slim X2 / Basal-IQ (Mmol/L), Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 1006419

Product 3

Affected products - t:slim/Control-IQ, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 1005611

Product 4

Affected products - t:slim/Control-IQ, Replacement, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 1005613

Risk / hazard

Tandem Diabetes Care, Inc. ("Tandem") has identified certain speaker versions that are in t slim X insulin pumps have a higher rate of malfunction 16 events due to a wiring issue within the speaker. Recall start date: August 18, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Tandem Diabetes Care, T:slim X Insulin Pumps.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

77883

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 8 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.