Synchron systems cholesterol (chol) reagent

Official notice

19 Nov 2021

Last checked

19 Nov 2021

Critical riskHealth Canada

An official Health Canada notice concerns Synchron systems cholesterol (chol) reagent. Beckman coulter has identified that Synchron Systems Cholesterol Reagent lot M106450 may fail with out of calibration range low (OCR low) and/or back to back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. The continued use of Synchron Systems cholesterol reagent lot M106450 may lead to: -Further failed QC and/or calibrations -Erroneous low Cholesterol patient results Recall start date: 2021-11-08 Published 2021-11-19.

What to do now

Contact the manufacturer if you require additional information

Recall number

63635

Official notice

2021-11-19

The hazard

Beckman coulter has identified that Synchron Systems Cholesterol Reagent lot M106450 may fail with out of calibration range low (OCR low) and/or back to back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. The continued use of Synchron Systems cholesterol reagent lot M106450 may lead to: -Further failed QC and/or calibrations -Erroneous low Cholesterol patient results Recall start date: 2021-11-08

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

M106450

Recall Numbers

63635

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT, Lot or serial number - M106450, Model or catalog number - 467825

Risk / hazard

Beckman coulter has identified that Synchron Systems Cholesterol Reagent lot M106450 may fail with out of calibration range low (OCR low) and/or back to back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. The continued use of Synchron Systems cholesterol reagent lot M106450 may lead to: -Further failed QC and/or calibrations -Erroneous low Cholesterol patient results Recall start date: 2021-11-08

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter Canada L.p, Synchron Cholesterol Reagent.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

63635

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 19 Nov 2021. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.