Statclave g4 chamber autoclaves

Official notice

10 Jul 2023

Last checked

10 Jul 2023

Medium riskHealth Canada

An official Health Canada notice concerns Statclave g4 chamber autoclaves. An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short. Recall Start Date: June 16, 2023 Published 2023-07-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73915

Official notice

2023-07-10

The hazard

An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short. Recall Start Date: June 16, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73915

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Statclave G4 Chamber Autoclaves, Lot or serial number - 710223E00008 710223E00017 710223E00010 710223E00016 710223E00007 710223E00014 710223E00011 710223E00013 710223E00012 710223E00009, Model or catalog number - G4-622301

Risk / hazard

An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short. Recall Start Date: June 16, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Scican, Statclave G4 Chamber Autoclaves.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

73915

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 10 Jul 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.