No image published for this recall

No image published for this recall

Spacelabs display cables for xprezzon® or qube bedside monitors

Official notice

17 Oct 2025

Last checked

17 Oct 2025

Medium riskHealth Canada

An official Health Canada notice concerns Spacelabs display cables for xprezzon® or qube bedside monitors. While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits. Published 2025-10-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78311

Official notice

2025-10-17

The hazard

While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78311

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen, Lot or serial number - 012-089X-XX (cable), Model or catalogue number - 94267

Product 2

Affected products - DVI Cables Used With Qube Compact Monitor, Lot or serial number - 012-089X-XX (cable), Model or catalogue number - 91390

Product 3

Affected products - DVI Cable Used With Bedside Monitor, Lot or serial number - 012-089X-XX (cable), Model or catalogue number - 91393

Risk / hazard

While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Spacelabs Healthcare, Spacelabs Display Cables For Xprezzon® Or Qube Bedside Monitors.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

78311

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 17 Oct 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.