Single use repositionable clip
Official notice
19 May 2023
Last checked
19 May 2023
An official Health Canada notice concerns Single use repositionable clip. Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure. The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. Hemostasis not achieved), prolonged procedure (e.g. Extended time to accomplish hemostasis), and unexpected medical interventions (e.g. Additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure. Recall start date: May 2, 2023 Published 2023-05-19.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73677
Official notice
2023-05-19
The hazard
The hazard
Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure. The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. Hemostasis not achieved), prolonged procedure (e.g. Extended time to accomplish hemostasis), and unexpected medical interventions (e.g. Additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure. Recall start date: May 2, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Olympus Medical, Single Use Repositionable Clip.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 19 May 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.