Single use repositionable clip

Official notice

19 May 2023

Last checked

19 May 2023

Critical riskHealth Canada

An official Health Canada notice concerns Single use repositionable clip. Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure. The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. Hemostasis not achieved), prolonged procedure (e.g. Extended time to accomplish hemostasis), and unexpected medical interventions (e.g. Additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure. Recall start date: May 2, 2023 Published 2023-05-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73677

Official notice

2023-05-19

The hazard

Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure. The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. Hemostasis not achieved), prolonged procedure (e.g. Extended time to accomplish hemostasis), and unexpected medical interventions (e.g. Additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure. Recall start date: May 2, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73677

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Single Use Repositionable Clip, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - HX-202LR HX-202UR

Risk / hazard

Olympus has become aware of increased complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure. The issues described above can lead to hazardous situations in which the clips cannot achieve their intended purpose, namely: (1) approximate a tissue opening or defect, (2) provide hemostasis to a bleeding vessel or mucosa, or (3) serve as a marker for subsequent location identification. Patient harms associated with these hazardous situations include: absence of treatment (e.g. Hemostasis not achieved), prolonged procedure (e.g. Extended time to accomplish hemostasis), and unexpected medical interventions (e.g. Additional clipping) being required to accomplish therapeutic purposes. Additionally, the issues described above can lead to mucosal injury/perforation, hemorrhage, and dehiscence of a previously approximated closure. Recall start date: May 2, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Olympus Medical, Single Use Repositionable Clip.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

critical

Recall details

Recall number

73677

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 19 May 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.