Reliance defibrillation leads

Official notice

29 Jul 2025

CriticalHealth Canada

An official Health Canada notice concerns Reliance defibrillation leads. Boston Scientific is initiating an advisory for ePTFE covered shock coil models from the RELIANCE family of defibrillation leads. The ePTFE material covering the shocking coils creates a matrix that, in some patients, promotes calcification, a biological wound response which is associated with gradual rising shock lead impedance (SLI) over time. Boston Scientific no longer manufactures or distributes ePTFE leads. Defibrillation leads currently distributed are non-ePTFE. Product is not being recalled. Recall start date: July 24, 2025 Published 2025-07-29.

What to do now

Healthcare professionals to conduct review of HVSI and determine if lead replacement is necessary.

Recall number

77782

Official notice

2025-07-29

The hazard

Boston Scientific is initiating an advisory for ePTFE covered shock coil models from the RELIANCE family of defibrillation leads. The ePTFE material covering the shocking coils creates a matrix that, in some patients, promotes calcification, a biological wound response which is associated with gradual rising shock lead impedance (SLI) over time. Boston Scientific no longer manufactures or distributes ePTFE leads. Defibrillation leads currently distributed are non-ePTFE. Product is not being recalled. Recall start date: July 24, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77782

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0696

Product 2

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0686

Product 3

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0692

Product 4

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0683

Product 5

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0693

Product 6

Affected products - RELIANCE 4-FRONT, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0695

Product 7

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0175

Product 8

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0184

Product 9

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0187

Product 10

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0165

Product 11

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0185

Product 12

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0295

Product 13

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0296

Product 14

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0186

Product 15

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0286

Product 16

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0164

Product 17

Affected products - ENDOTAK RELIANCE G, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0167

Product 18

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0181

Product 19

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0292

Product 20

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0161

Product 21

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0282

Product 22

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0283

Product 23

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0170

Product 24

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0293

Product 25

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0171

Product 26

Affected products - ENDOTAK RELIANCE SG, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 0180

Risk / hazard

Boston Scientific is initiating an advisory for ePTFE covered shock coil models from the RELIANCE family of defibrillation leads. The ePTFE material covering the shocking coils creates a matrix that, in some patients, promotes calcification, a biological wound response which is associated with gradual rising shock lead impedance (SLI) over time. Boston Scientific no longer manufactures or distributes ePTFE leads. Defibrillation leads currently distributed are non-ePTFE. Product is not being recalled. Recall start date: July 24, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Boston Scientific, Reliance Defibrillation Leads.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

critical

Recall details

Recall number

77782

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Medium

Faradrive™ steerable sheath

Health Canada19 Jun 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Icefx™ cryoablation console

Health Canada11 Sept 2025

Brand match

Medium risk

Markets

CA
HEALTH
High

Accolade™ and visionist™ crt-p pacemakers

Health Canada22 Aug 2025

Brand match

High risk

Markets

CA
HEALTH
Medium

Encore™ 26 inflation device, encore™ 26 advantage kits, nephromax™ kits and uromax ultra™ kits

Health Canada14 Aug 2025

Brand match

Medium risk

Markets

CA
HEALTH
High

Watchman™ access systems

Health Canada30 Jul 2025

Brand match

High risk

Markets

CA
HEALTH
High

Carotid wallstent™ monorail

Health Canada11 Jul 2025

Brand match

High risk

Markets

CA
HEALTH
Medium

Avvigo™+ multi-modality guidance system

Health Canada18 Dec 2024

Brand match

Medium risk

Markets

CA
HEALTH
Critical

Acurate neo2 and acurate prime

Health Canada2 Dec 2024

Brand match

Critical risk

Markets

CA
HEALTH
High

Polarx fit short and long tip cryoablation balloon catheter

Health Canada15 Oct 2024

Brand match

High risk

Markets

CA
HEALTH
High

Accolade™ mri pacemaker

Health Canada24 Sept 2024

Brand match

High risk

Markets

CA
HEALTH
Medium

Expo angiographic catheters

Health Canada3 Apr 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Isleeve™ expandable introducer set

Health Canada29 Feb 2024

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 29 Jul 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.