No image published for this recall
Radiant warmer and birthing room warmer
Official notice
17 May 2023
Last checked
17 May 2023
An official Health Canada notice concerns Radiant warmer and birthing room warmer. During the course of global market surveillance activities, the manufaturer has become aware of cases in which the resuscitaire infant radiant warmer with the optional scale could display inaccurate weight values. There is a potential for the mattress cover material to slightly interfere with the corner brackets and/or side panels of the bassinet. When this condition is present, the scale may not "zero" correctly during a "weigh" function, which may lead to an inaccurate weight measurement of the infant. Recall start date: April 28, 2023 Published 2023-05-17.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73686
Official notice
2023-05-17
The hazard
The hazard
During the course of global market surveillance activities, the manufaturer has become aware of cases in which the resuscitaire infant radiant warmer with the optional scale could display inaccurate weight values. There is a potential for the mattress cover material to slightly interfere with the corner brackets and/or side panels of the bassinet. When this condition is present, the scale may not "zero" correctly during a "weigh" function, which may lead to an inaccurate weight measurement of the infant. Recall start date: April 28, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Draeger Medical, Radiant Warmer And Birthing Room Warmer.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Infinity® centralstation wide

Infinity m500 docking station

Infinity® central station

Bililux

Infinity centralstation wide

Intramedullary bone saw blade assembly

Aisys, avance, and aespire anesthesia machines

Glidesheath slender® introducer sheath
Image: Health Canada recall noticeImportation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Bd maxplus ™ connector minibore extension set

Corneat kpro

Artis icono ceiling
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 17 May 2023. The next one won't wait for you to search.