No image published for this recall

No image published for this recall

Power express centrifuge

Official notice

1 Aug 2024

Last checked

1 Aug 2024

Medium riskHealth Canada

An official Health Canada notice concerns Power express centrifuge. Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. Recall start date: July 22, 2024 Published 2024-08-01.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75888

Official notice

2024-08-01

The hazard

Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. Recall start date: July 22, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75888

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Power Express Centrifuge, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - B36365 B36366

Risk / hazard

Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. Recall start date: July 22, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Power Express Centrifuge.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75888

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 1 Aug 2024. The next one won't wait for you to search.