Plum infusion systems
Official notice
18 Apr 2023
Last checked
18 Apr 2023
An official Health Canada notice concerns Plum infusion systems. If a Plum 360 or Plum A+ Infusion System is running on battery power, a low battery and depleted battery alarm should typically activate when thirty minutes and three minutes, respectively, of estimated battery runtime remain. In addition, when the pump detects a loss of battery capacity, the pump will display a message to replace the battery. On a Plum 360 pump, the screen will display "Keep plugged into AC! Service battery / replace pump" and Plum A+ pump will display "Warning: replace battery." Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected. If an affected battery is not replaced when the pump displays the message to replace the battery, there may be less than thirty minutes of battery runtime after the low battery alarm resulting in the depleted battery alarm occurring sooner than expected. At this point, the pump stops delivery, alarms depleted battery for three minutes, and the pump turns off. Please note: This recall may also affect spare batteries from CSB, manufactured before January 1, 2023. If the first two characters on the label are 22 or lower, the battery is potentially affected. Recall start date: March 12, 2023 Published 2023-04-18.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73535
Official notice
2023-04-18
The hazard
The hazard
If a Plum 360 or Plum A+ Infusion System is running on battery power, a low battery and depleted battery alarm should typically activate when thirty minutes and three minutes, respectively, of estimated battery runtime remain. In addition, when the pump detects a loss of battery capacity, the pump will display a message to replace the battery. On a Plum 360 pump, the screen will display "Keep plugged into AC! Service battery / replace pump" and Plum A+ pump will display "Warning: replace battery." Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected. If an affected battery is not replaced when the pump displays the message to replace the battery, there may be less than thirty minutes of battery runtime after the low battery alarm resulting in the depleted battery alarm occurring sooner than expected. At this point, the pump stops delivery, alarms depleted battery for three minutes, and the pump turns off. Please note: This recall may also affect spare batteries from CSB, manufactured before January 1, 2023. If the first two characters on the label are 22 or lower, the battery is potentially affected. Recall start date: March 12, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Icu Medical, Plum Infusion.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Plum™ precision IV pumps

Plum solo and plum duo precision IV pumps

Hotline IV fluid warmer system & accessories

Swabsleeve™ and swabcap™

Icu medical burette set (clave, shut-off)

Cadd™ medication cassette reservoirs

Tego connector

Icu medical burette set (clave, shut-off)

Plum infusion systems

Plum a+3 infusion pumps

Plum 360 infusion system with mednet

Microclave clear sets
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 18 Apr 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.