Pico50 arterial blood sampler

Official notice

25 Apr 2022

Last checked

25 Apr 2022

Medium riskHealth Canada

An official Health Canada notice concerns Pico50 arterial blood sampler. Radiometer has been performing integrity testing of the Pico50 arterial blood sampler packaging material. Integrity studies included sterility testing to show that the integrity of the sterile barrier system is intact. In the 24 month real time study, 1 out of 40 samples did not pass the sterility test acceptance criteria. This indicates that the sterile barrier system is compromised after 2 years storage and maintaining the product sterile throughout the shelf life cannot be showed. Recall start date: Apr 13, 2022 Published 2022-04-25.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64108

Official notice

2022-04-25

The hazard

Radiometer has been performing integrity testing of the Pico50 arterial blood sampler packaging material. Integrity studies included sterility testing to show that the integrity of the sterile barrier system is intact. In the 24 month real time study, 1 out of 40 samples did not pass the sterility test acceptance criteria. This indicates that the sterile barrier system is compromised after 2 years storage and maintaining the product sterile throughout the shelf life cannot be showed. Recall start date: Apr 13, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64108

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Pico50 Arterial Blood Sampler, Lot or Serial Number - All lots., Model or catalog number - 956-552

Risk / hazard

Radiometer has been performing integrity testing of the Pico50 arterial blood sampler packaging material. Integrity studies included sterility testing to show that the integrity of the sterile barrier system is intact. In the 24 month real time study, 1 out of 40 samples did not pass the sterility test acceptance criteria. This indicates that the sterile barrier system is compromised after 2 years storage and maintaining the product sterile throughout the shelf life cannot be showed. Recall start date: Apr 13, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Radiometer Medical Aps, Pico50 Arterial Blood Sampler.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64108

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 25 Apr 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.