No image published for this recall
Phoroptor vrx digital refraction system
Official notice
8 Sept 2023
Last checked
8 Sept 2023
An official Health Canada notice concerns Phoroptor vrx digital refraction system. Some Phoroptor VRx devices may have an assembly defect that causes the device to detach from the bracket that connects it to an ophthalmic stand. A device that detaches from a stand can fall on to a patient or operator causing an injury. The defect was identified via a customer complaint received on April 3, 2023. The subsequent complaint investigation showed that other units built in December 2021 and January 2022 could have the same defect. Recall Start Date: August 11, 2023 Published 2023-09-08.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74241
Official notice
2023-09-08
The hazard
The hazard
Some Phoroptor VRx devices may have an assembly defect that causes the device to detach from the bracket that connects it to an ophthalmic stand. A device that detaches from a stand can fall on to a patient or operator causing an injury. The defect was identified via a customer complaint received on April 3, 2023. The subsequent complaint investigation showed that other units built in December 2021 and January 2022 could have the same defect. Recall Start Date: August 11, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: REI, Phoroptor Vrx Digital Refraction System.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 8 Sept 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.