No image published for this recall

No image published for this recall

Philips sense xl torso (1.5t and 3.0t) coils

Official notice

13 Jun 2024

Last checked

13 Jun 2024

High riskHealth Canada

An official Health Canada notice concerns Philips sense xl torso (1.5t and 3.0t) coils. Philips has been made aware through customer complaints of a potential safety issue with specific SENSE XL Torso (1.5T and 3.0T) coils that could pose a risk for patients. Philips has identified an issue in the SENSE XL Torso (1.5T and 3.0T) coils which may result in localized heating during the scan, resulting in potential harm to patients. If the patient is exposed to localized heating there is a potential for a heating sensation and/or 1st, 2nd, or 3rd degree burns around the area covered by the coil. Recall start date: June 6, 2024 Published 2024-06-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75658

Official notice

2024-06-13

The hazard

Philips has been made aware through customer complaints of a potential safety issue with specific SENSE XL Torso (1.5T and 3.0T) coils that could pose a risk for patients. Philips has identified an issue in the SENSE XL Torso (1.5T and 3.0T) coils which may result in localized heating during the scan, resulting in potential harm to patients. If the patient is exposed to localized heating there is a potential for a heating sensation and/or 1st, 2nd, or 3rd degree burns around the area covered by the coil. Recall start date: June 6, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75658

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Achieva 3.0T MRI System - Sense Torso Coil, Lot or serial number - Not applicable., Model or catalog number - 4535 673 94942 4535-673-9494X

Product 2

Affected products - Intera 1.5T MRI System - Sense XL Torso Coil, Lot or serial number - Not applicable., Model or catalog number - 453567141882 453567141883

Product 3

Affected products - Achieva 1.5T MRI System - Sense Torso Coil, Lot or serial number - Not applicable., Model or catalog number - 453567141882 453567141883 4535 675 02881

Product 4

Affected products - Achieva XR System - Sense Coil, Lot or serial number - Not applicable., Model or catalog number - 453567141882 453567141883 453567394942

Risk / hazard

Philips has been made aware through customer complaints of a potential safety issue with specific SENSE XL Torso (1.5T and 3.0T) coils that could pose a risk for patients. Philips has identified an issue in the SENSE XL Torso (1.5T and 3.0T) coils which may result in localized heating during the scan, resulting in potential harm to patients. If the patient is exposed to localized heating there is a potential for a heating sensation and/or 1st, 2nd, or 3rd degree burns around the area covered by the coil. Recall start date: June 6, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

75658

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 13 Jun 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.