No image published for this recall

No image published for this recall

Philips bipap and humidifiers

Official notice

10 Jul 2023

Last checked

10 Jul 2023

Medium riskHealth Canada

An official Health Canada notice concerns Philips bipap and humidifiers. There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device components are unknown. Full validation requires biological evaluation after cleaning cycles per ISO 10993-1 and testing on expected worse-case clinical use per ISO 18562-1. The maximum number of Keredusy cycles was also not identified within the user manuals. ISO 17664-1 does allow for the manufacturer's IFU to direct the processor to refer to the disinfecting agent manufacturer's IFU with reference to concentration, temperature, and contact time, but these parameters were not defined nor documented within the IFU. The IFU only directed the user to Philips customer service. Recall Start Date: June 30, 2023 Published 2023-07-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73914

Official notice

2023-07-10

The hazard

There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device components are unknown. Full validation requires biological evaluation after cleaning cycles per ISO 10993-1 and testing on expected worse-case clinical use per ISO 18562-1. The maximum number of Keredusy cycles was also not identified within the user manuals. ISO 17664-1 does allow for the manufacturer's IFU to direct the processor to refer to the disinfecting agent manufacturer's IFU with reference to concentration, temperature, and contact time, but these parameters were not defined nor documented within the IFU. The IFU only directed the user to Philips customer service. Recall Start Date: June 30, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73914

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - BiPAP A40 Pro, CA, Lot or serial number - All lots., Model or catalog number - CAX3100S12

Product 2

Affected products - BiPAP A40, Canada, Core Package, Lot or serial number - All lots., Model or catalog number - 1111174

Product 3

Affected products - BiPAP A30 System - Ventilator & System One Aseries Heated Humidifier, Lot or serial number - All lots., Model or catalog number - 1111148

Product 4

Affected products - BiPAP A40 EFL, Lot or serial number - All lots., Model or catalog number - INX3T

Product 5

Affected products - System One Heated Humidifier, Lot or serial number - All lots., Model or catalog number - IN6T INX3T IN6TFLG IN6H INX3H IN6HFLG

Product 6

Affected products - BiPAP A40 Pro, CA W/ Heated Tube Humidifier, CA, Lot or serial number - All Lots., Model or catalog number - CAX3100T12

Product 7

Affected products - System One A-Series Heated Humidifier, Lot or serial number - All Lots., Model or catalog number - 1111552 1077107 1111555

Risk / hazard

There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device components are unknown. Full validation requires biological evaluation after cleaning cycles per ISO 10993-1 and testing on expected worse-case clinical use per ISO 18562-1. The maximum number of Keredusy cycles was also not identified within the user manuals. ISO 17664-1 does allow for the manufacturer's IFU to direct the processor to refer to the disinfecting agent manufacturer's IFU with reference to concentration, temperature, and contact time, but these parameters were not defined nor documented within the IFU. The IFU only directed the user to Philips customer service. Recall Start Date: June 30, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

73914

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 10 Jul 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.