Patient information center ix

Official notice

14 Mar 2024

Last checked

14 Mar 2024

Medium riskHealth Canada

An official Health Canada notice concerns Patient information center ix. It was found that the care assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when ios device screens were locked with a passcode. The issue occurred on devices running IOS 15, 16 and 17 as well as multiple versions of iPhones. The care assist mobile application is not intended to replace the patient's primary alarming system. If users rely on the care assist application as their primary event notification alarming system, the failure of custom ringtones could result in a delay in treatment for patients experiencing a change in clinical condition. Recall Start Date: March 6, 2024 Published 2024-03-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75241

Official notice

2024-03-14

The hazard

It was found that the care assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when ios device screens were locked with a passcode. The issue occurred on devices running IOS 15, 16 and 17 as well as multiple versions of iPhones. The care assist mobile application is not intended to replace the patient's primary alarming system. If users rely on the care assist application as their primary event notification alarming system, the failure of custom ringtones could result in a delay in treatment for patients experiencing a change in clinical condition. Recall Start Date: March 6, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75241

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Patient Information Center iX, Lot or serial number - Not applicable., Model or catalog number - VERSION 4.0 VERSION 4.1

Risk / hazard

It was found that the care assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when ios device screens were locked with a passcode. The issue occurred on devices running IOS 15, 16 and 17 as well as multiple versions of iPhones. The care assist mobile application is not intended to replace the patient's primary alarming system. If users rely on the care assist application as their primary event notification alarming system, the failure of custom ringtones could result in a delay in treatment for patients experiencing a change in clinical condition. Recall Start Date: March 6, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips, Patient Information Center Ix.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75241

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 Mar 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.