Orise proknife procedure and gel syringe kits

Official notice

26 Oct 2022

Last checked

26 Oct 2022

Medium riskHealth Canada

An official Health Canada notice concerns Orise proknife procedure and gel syringe kits. Boston Scientific recently became aware of events associated with foreign body reaction which presented as mass formations from remnant Orise gel post procedure, prompting unnecessary surgical intervention at a rate of 0.0019% of units sold (based on known events and total sales to date). There have been no reported deaths associated with these events. As a result of these events, updates will be made to the product IFUs. The purpose of the updates is to (a) raise awareness that remnant Orise gel may elicit a foreign body reaction with granuloma and multinucleate giant cells that can physically appear as mass formations and submucosal distortions, (b) provide new warnings and precautions, and (c) reinforce the existing procedural instructions and the need to document the use of Orise gel during the procedure. Boston scientific is not removing any Orise gel devices from the field; devices remain available for use. Recall start date: October 11, 2022 Published 2022-10-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72078

Official notice

2022-10-26

The hazard

Boston Scientific recently became aware of events associated with foreign body reaction which presented as mass formations from remnant Orise gel post procedure, prompting unnecessary surgical intervention at a rate of 0.0019% of units sold (based on known events and total sales to date). There have been no reported deaths associated with these events. As a result of these events, updates will be made to the product IFUs. The purpose of the updates is to (a) raise awareness that remnant Orise gel may elicit a foreign body reaction with granuloma and multinucleate giant cells that can physically appear as mass formations and submucosal distortions, (b) provide new warnings and precautions, and (c) reinforce the existing procedural instructions and the need to document the use of Orise gel during the procedure. Boston scientific is not removing any Orise gel devices from the field; devices remain available for use. Recall start date: October 11, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72078

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Orise Proknife Procedure Kit, Lot or serial number - All lots., Model or catalog number - M00519400 M00519380 M00519390

Product 2

Affected products - Orise Gel Syringe Single Pack, Lot or serial number - All lots., Model or catalog number - M00519231 M00519221

Product 3

Affected products - Orise Gel Syringe Twin Pack, Lot or serial number - All lots., Model or catalog number - M00519210 M00519211 M00519201

Risk / hazard

Boston Scientific recently became aware of events associated with foreign body reaction which presented as mass formations from remnant Orise gel post procedure, prompting unnecessary surgical intervention at a rate of 0.0019% of units sold (based on known events and total sales to date). There have been no reported deaths associated with these events. As a result of these events, updates will be made to the product IFUs. The purpose of the updates is to (a) raise awareness that remnant Orise gel may elicit a foreign body reaction with granuloma and multinucleate giant cells that can physically appear as mass formations and submucosal distortions, (b) provide new warnings and precautions, and (c) reinforce the existing procedural instructions and the need to document the use of Orise gel during the procedure. Boston scientific is not removing any Orise gel devices from the field; devices remain available for use. Recall start date: October 11, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Boston Scientific, Orise Proknife Procedure And Gel Syringe Kits.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

72078

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 26 Oct 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.