Novum iq large volume pump

Official notice

6 Feb 2023

Last checked

6 Feb 2023

High riskHealth Canada

An official Health Canada notice concerns Novum iq large volume pump. Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval. Baxter Corporation is issuing a follow-up communication to notify users that all of the sets on the Novum IQ LVP compatible sets list are acceptable to use at rates below 1mL/hr except for the product code 1W5000, which will be removed from the Novum IQ LVP list of compatible sets. Recall start date: September 15, 2022 Published 2023-02-06.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64580

Official notice

2023-02-06

The hazard

Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval. Baxter Corporation is issuing a follow-up communication to notify users that all of the sets on the Novum IQ LVP compatible sets list are acceptable to use at rates below 1mL/hr except for the product code 1W5000, which will be removed from the Novum IQ LVP list of compatible sets. Recall start date: September 15, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64580

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Novum IQ Large Volume Pump, Lot or serial number - All lots., Model or catalog number - 40700BAX

Risk / hazard

Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval. Baxter Corporation is issuing a follow-up communication to notify users that all of the sets on the Novum IQ LVP compatible sets list are acceptable to use at rates below 1mL/hr except for the product code 1W5000, which will be removed from the Novum IQ LVP list of compatible sets. Recall start date: September 15, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Baxter Healthcare, Novum Iq Large Volume Pump.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

64580

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 6 Feb 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.