Neo-tee t-piece resuscitator

Official notice

24 Dec 2024

Last checked

24 Dec 2024

High riskHealth Canada

An official Health Canada notice concerns Neo-tee t-piece resuscitator. To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller. Recall start date: December 5, 2024 Published 2024-12-24.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76755

Official notice

2024-12-24

The hazard

To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller. Recall start date: December 5, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76755

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Neo-Tee T-Piece Resuscitator, Lot or serial number - 2426450809, Model or catalogue number - 10-50809

Risk / hazard

To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller. Recall start date: December 5, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Mercury Medical, Neo-tee T-piece Resuscitator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

76755

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 24 Dec 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.