An official Health Canada notice concerns Monnal™ t60 ventilator. The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation. Published 2025-10-22.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78347

Official notice

2025-10-22

The hazard

The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78347

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Monnal™ T60 Ventilator, Lot or serial number - Not applicable., Model or catalogue number - KA017115

Product 2

Affected products - Expiratory Valve Monnal EVA V2, Lot or serial number - Not applicable., Model or catalogue number - KY732991

Risk / hazard

The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Air Liquide Medical, Monnal™ T60 Ventilator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

78347

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 22 Oct 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.