No image published for this recall

No image published for this recall

Minimed 630g and 670g

Official notice

3 Oct 2022

Last checked

3 Oct 2022

Critical riskHealth Canada

An official Health Canada notice concerns Minimed 630g and 670g. Communication between the components of the MINIMED 600 Series Pump System could be compromised by unauthorized access. This can lead to too much or too little insulin being delivered through delivery of unintended insulin bolus or because insulin delivery is slowed or stopped. Recall start date: September 22, 2022 Published 2022-10-03.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64595

Official notice

2022-10-03

The hazard

Communication between the components of the MINIMED 600 Series Pump System could be compromised by unauthorized access. This can lead to too much or too little insulin being delivered through delivery of unintended insulin bolus or because insulin delivery is slowed or stopped. Recall start date: September 22, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64595

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Minimed 630g, Lot or serial number - All lots., Model or catalogue number - MMT-1714K MMT-1714KRP MMT-1714

Product 2

Affected products - Minimed 670g, Lot or serial number - All lots., Model or catalogue number - MMT-1782K MMT-1782 MMT-1782KRP MMT-1782CRP

Product 3

Affected products - Minimed 670g Kit, Lot or serial number - All lots., Model or catalogue number - MMT-1762KCN MMT-1762CRP

Risk / hazard

Communication between the components of the MINIMED 600 Series Pump System could be compromised by unauthorized access. This can lead to too much or too little insulin being delivered through delivery of unintended insulin bolus or because insulin delivery is slowed or stopped. Recall start date: September 22, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: MINI, Minimed 630g And 670g.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

64595

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 3 Oct 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.