No image published for this recall

No image published for this recall

Marquee disposable core biopsy instrument kit

Official notice

9 Oct 2024

Last checked

9 Oct 2024

Medium riskHealth Canada

An official Health Canada notice concerns Marquee disposable core biopsy instrument kit. Based on customer complaints, Bard Peripheral Vascular, Inc. Has identified that specific lots of the Marquee™ Disposable Core Biopsy Instruments may have been potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device. Recall Start Date: September 30, 2024 Published 2024-10-09.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76311

Official notice

2024-10-09

The hazard

Based on customer complaints, Bard Peripheral Vascular, Inc. Has identified that specific lots of the Marquee™ Disposable Core Biopsy Instruments may have been potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device. Recall Start Date: September 30, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76311

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Marquee Disposable Core Biopsy Instrument, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - MQ1210 MQ1213

Product 2

Affected products - Marquee Disposable Core Biopsy Instrument Kit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - MQK1210 MQK1213

Risk / hazard

Based on customer complaints, Bard Peripheral Vascular, Inc. Has identified that specific lots of the Marquee™ Disposable Core Biopsy Instruments may have been potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device. Recall Start Date: September 30, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Bard Peripheral Vascular, Marquee Disposable Core Biopsy Instrument Kit.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

76311

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 9 Oct 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.