No image published for this recall

No image published for this recall

Low prof lckng screw f im nail, sterile

Official notice

8 Oct 2025

Last checked

8 Oct 2025

Medium riskHealth Canada

An official Health Canada notice concerns Low prof lckng screw f im nail, sterile. The affected product lots are being recalled because, although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm. Published 2025-10-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

78256

Official notice

2025-10-08

The hazard

The affected product lots are being recalled because, although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

78256

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Low Prof Lckng Screw F/Im Nail, Sterile, Lot or serial number - 04.045.346TS, Model or catalogue number - 31756P7

Product 2

Affected products - Low Prof Lckng Screw F/Im Nail, Sterile, Lot or serial number - 04.045.342TS, Model or catalogue number - 31756P9

Risk / hazard

The affected product lots are being recalled because, although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Synthes, Low Prof Lckng Screw F Im Nail Sterile.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

78256

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 8 Oct 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.