No image published for this recall
Liko m220 m230 mobile lift
Official notice
24 Jan 2022
Last checked
24 Jan 2022
An official Health Canada notice concerns Liko m220 m230 mobile lift. Hillrom received four (4) complaints between 29-Sep-2020 and 09-Apr-2021 of the Slingbar 450 becoming detached from the upper bracket of the Slingbar attachment on Liko M230 Mobile Lifts. The Slingbar is attached to the lifting arm through two brackets. These two brackets are attached to each other by a bolt and a sling plate that allows for rotation of the sling bar in one axis relative to the upper bracket. The investigation identified the root cause of the failure to be that only one (1) instead of two (2) plastic bushings had been installed during production. The absence of the lower plastic bushings can lead to the center bolt wearing on the sling plate linkage mechanism. Excessive wear can result in inadequate attachment of the Slingbar to the lift, causing it to become detached and resulting in the patient being dropped from the lift. Recall start date: Jan 13, 2022 Published 2022-01-24.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63860
Official notice
2022-01-24
The hazard
The hazard
Hillrom received four (4) complaints between 29-Sep-2020 and 09-Apr-2021 of the Slingbar 450 becoming detached from the upper bracket of the Slingbar attachment on Liko M230 Mobile Lifts. The Slingbar is attached to the lifting arm through two brackets. These two brackets are attached to each other by a bolt and a sling plate that allows for rotation of the sling bar in one axis relative to the upper bracket. The investigation identified the root cause of the failure to be that only one (1) instead of two (2) plastic bushings had been installed during production. The absence of the lower plastic bushings can lead to the center bolt wearing on the sling plate linkage mechanism. Excessive wear can result in inadequate attachment of the Slingbar to the lift, causing it to become detached and resulting in the patient being dropped from the lift. Recall start date: Jan 13, 2022
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Hillrom Luleå, Liko M220 M230 Mobile Lift.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 24 Jan 2022. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.