An official Health Canada notice concerns Liberator stationary liquid oxygen device. There were inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle. This can lead to premature inner vessel failure. The failure mode would be potential loss of vacuum, which could cause either an over-delivery of oxygen or the outer vessel to frost. Increased flow of oxygen from the relief valves could lead to an increased fire risk. Recall start date: October 11, 2022 Published 2022-11-09.

What to do now

Contact the manufacturer if you require additional information.

Recall number

72172

Official notice

2022-11-09

The hazard

There were inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle. This can lead to premature inner vessel failure. The failure mode would be potential loss of vacuum, which could cause either an over-delivery of oxygen or the outer vessel to frost. Increased flow of oxygen from the relief valves could lead to an increased fire risk. Recall start date: October 11, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

72172

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Liberator Stationary Liquid Oxygen Device, Lot or serial number - CBB3022320074 CBB3022310852 CBB3022310843 CBB3022310827 CBB3022310823 CBB3022310842 CBB3022310845 CBB3022320077 CBB3022310792 CBB3022310844, Model or catalog number - 13261699 13256195

Risk / hazard

There were inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle. This can lead to premature inner vessel failure. The failure mode would be potential loss of vacuum, which could cause either an over-delivery of oxygen or the outer vessel to frost. Increased flow of oxygen from the relief valves could lead to an increased fire risk. Recall start date: October 11, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Caire, Liberator Stationary Liquid Oxygen Device.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

72172

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 9 Nov 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.