No image published for this recall

No image published for this recall

Invacare tdx sp2 with linx gyro

Official notice

26 Nov 2021

Last checked

26 Nov 2021

Medium riskHealth Canada

An official Health Canada notice concerns Invacare tdx sp2 with linx gyro. Power wheelchairs with a linx Gyro component running firmware version 6.1.2 Can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile. Recall start date: 2021-11-12 Published 2021-11-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63656

Official notice

2021-11-26

The hazard

Power wheelchairs with a linx Gyro component running firmware version 6.1.2 Can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile. Recall start date: 2021-11-12

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63656

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - INVACARE TDX SP2 WITH LINX GYRO, Lot or serial number - Not applicable., Model or catalog number - TDXSP2 TDXSP2-CG TDXSP2-MCG TDXSP2HD TDXSP2HD-CG TDXSP2HD-MCG TDXSP2V TDXSP2V-HD TDXSP2X-CG TDXSP2X-MCG

Risk / hazard

Power wheelchairs with a linx Gyro component running firmware version 6.1.2 Can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile. Recall start date: 2021-11-12

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Invacare, Invacare Tdx Sp2 With Linx Gyro.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

63656

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 26 Nov 2021. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.