Intelepacs (image fusion module)

Official notice

19 Feb 2025

Last checked

19 Feb 2025

Critical riskHealth Canada

An official Health Canada notice concerns Intelepacs (image fusion module). An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis. Recall start date: February 4, 2025 Published 2025-02-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76978

Official notice

2025-02-19

The hazard

An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis. Recall start date: February 4, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76978

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - IntelePACS (Image Fusion Module), Lot or serial number - Software version 5-6-1-P23 and later, Model or catalog number - INTELEPACS

Risk / hazard

An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis. Recall start date: February 4, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Intelerad Medical, Intelepacs.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

76978

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 19 Feb 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.