Medium riskHealth Canada

An official Health Canada notice concerns Intelepacs. There is a possibility that a patient's final report sent out of Intelepacs contains the preliminary text instead of the final report text. This issue can manifest itself only: - In specific pacs-driven workflow configurations (e.g. Multi-method reporting with transcriptionist worklist) where the signing step is handled by Intelepacs, - Specifically when a network connectivity problem occurs between the server the user is connected to and the primary RIS gateway server (rig) of the system, and - When the rig servers are not fully synchronized. Recall start date: Apr 25, 2022 Published 2022-05-03.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64138

Official notice

2022-05-03

The hazard

There is a possibility that a patient's final report sent out of Intelepacs contains the preliminary text instead of the final report text. This issue can manifest itself only: - In specific pacs-driven workflow configurations (e.g. Multi-method reporting with transcriptionist worklist) where the signing step is handled by Intelepacs, - Specifically when a network connectivity problem occurs between the server the user is connected to and the primary RIS gateway server (rig) of the system, and - When the rig servers are not fully synchronized. Recall start date: Apr 25, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64138

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Intelepacs, Lot or Serial Number - Version 4-10-1 and above, Model or catalog number - INTELEPACS

Risk / hazard

There is a possibility that a patient's final report sent out of Intelepacs contains the preliminary text instead of the final report text. This issue can manifest itself only: - In specific pacs-driven workflow configurations (e.g. Multi-method reporting with transcriptionist worklist) where the signing step is handled by Intelepacs, - Specifically when a network connectivity problem occurs between the server the user is connected to and the primary RIS gateway server (rig) of the system, and - When the rig servers are not fully synchronized. Recall start date: Apr 25, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Intelerad Medical, Intelepacs.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64138

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 3 May 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.