No image published for this recall
Ingenia ambition x
Official notice
18 Oct 2023
Last checked
18 Oct 2023
An official Health Canada notice concerns Ingenia ambition x. The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. During decommissioning or moving activities. The initial design of the hoisting blocks on top of the magnet did not meet the required mechanical specification. The diameter to meet the required bolt‐on interface surface for the swivel eye bolts, used as lifting hardware, was too small. A hazardous situation could occur, only when the magnet is lifted by crane, via attachments to these lifting blocks. Large angles between the individual slings connected to the lifting blocks of the magnet can result in increased stresses in these blocks. When the lifting interface would fail, a person beneath of near the magnet that is hoisted could be hit. No incidents have been reported. Recall Start Date: October 3, 2023 Published 2023-10-18.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74434
Official notice
2023-10-18
The hazard
The hazard
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. During decommissioning or moving activities. The initial design of the hoisting blocks on top of the magnet did not meet the required mechanical specification. The diameter to meet the required bolt‐on interface surface for the swivel eye bolts, used as lifting hardware, was too small. A hazardous situation could occur, only when the magnet is lifted by crane, via attachments to these lifting blocks. Large angles between the individual slings connected to the lifting blocks of the magnet can result in increased stresses in these blocks. When the lifting interface would fail, a person beneath of near the magnet that is hoisted could be hit. No incidents have been reported. Recall Start Date: October 3, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Philips, Ingenia Ambition X.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 18 Oct 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.