No image published for this recall

No image published for this recall

Ingenia ambition x

Official notice

18 Oct 2023

Last checked

18 Oct 2023

Medium riskHealth Canada

An official Health Canada notice concerns Ingenia ambition x. The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. During decommissioning or moving activities. The initial design of the hoisting blocks on top of the magnet did not meet the required mechanical specification. The diameter to meet the required bolt‐on interface surface for the swivel eye bolts, used as lifting hardware, was too small. A hazardous situation could occur, only when the magnet is lifted by crane, via attachments to these lifting blocks. Large angles between the individual slings connected to the lifting blocks of the magnet can result in increased stresses in these blocks. When the lifting interface would fail, a person beneath of near the magnet that is hoisted could be hit. No incidents have been reported. Recall Start Date: October 3, 2023 Published 2023-10-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74434

Official notice

2023-10-18

The hazard

The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. During decommissioning or moving activities. The initial design of the hoisting blocks on top of the magnet did not meet the required mechanical specification. The diameter to meet the required bolt‐on interface surface for the swivel eye bolts, used as lifting hardware, was too small. A hazardous situation could occur, only when the magnet is lifted by crane, via attachments to these lifting blocks. Large angles between the individual slings connected to the lifting blocks of the magnet can result in increased stresses in these blocks. When the lifting interface would fail, a person beneath of near the magnet that is hoisted could be hit. No incidents have been reported. Recall Start Date: October 3, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74434

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Ingenia Ambition X, Lot or serial number - Not applicable., Model or catalog number - 782109 781356

Risk / hazard

The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. During decommissioning or moving activities. The initial design of the hoisting blocks on top of the magnet did not meet the required mechanical specification. The diameter to meet the required bolt‐on interface surface for the swivel eye bolts, used as lifting hardware, was too small. A hazardous situation could occur, only when the magnet is lifted by crane, via attachments to these lifting blocks. Large angles between the individual slings connected to the lifting blocks of the magnet can result in increased stresses in these blocks. When the lifting interface would fail, a person beneath of near the magnet that is hoisted could be hit. No incidents have been reported. Recall Start Date: October 3, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips, Ingenia Ambition X.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74434

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.