Idylla instrument

Official notice

30 Dec 2024

Last checked

30 Dec 2024

Medium riskHealth Canada

An official Health Canada notice concerns Idylla instrument. Biocartis received a complaint on several Idylla MSI test false positive results generated on a single instrument. As part of the investigation, Biocartis conducted a review of the Idylla MSI test performance in the field. A review of all connected runs from 2019 to date identified suspected false positive Idylla MSI test runs obtained by customers in Germany, Italy, Poland, Estonia, Cyprus, and Thailand. Recall start date: December 9, 2024 Published 2024-12-30.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76770

Official notice

2024-12-30

The hazard

Biocartis received a complaint on several Idylla MSI test false positive results generated on a single instrument. As part of the investigation, Biocartis conducted a review of the Idylla MSI test performance in the field. A review of all connected runs from 2019 to date identified suspected false positive Idylla MSI test runs obtained by customers in Germany, Italy, Poland, Estonia, Cyprus, and Thailand. Recall start date: December 9, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

76770

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Idylla Instrument, Lot or serial number - SER2149 SER3468 SER2361 SER3460 SER3302 SER3321 SER3322 SER2501, Model or catalogue number - P0010

Risk / hazard

Biocartis received a complaint on several Idylla MSI test false positive results generated on a single instrument. As part of the investigation, Biocartis conducted a review of the Idylla MSI test performance in the field. A review of all connected runs from 2019 to date identified suspected false positive Idylla MSI test runs obtained by customers in Germany, Italy, Poland, Estonia, Cyprus, and Thailand. Recall start date: December 9, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Biocartis, Idylla Instrument.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

76770

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 30 Dec 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.