Horizon x-ray bone densitometer (dxa)
Official notice
21 Jun 2024
Last checked
21 Jun 2024
An official Health Canada notice concerns Horizon x-ray bone densitometer (dxa). During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024 Published 2024-06-21.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75719
Official notice
2024-06-21
The hazard
The hazard
During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Hologic, Horizon X-ray Bone Densitometer.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Selenia dimensions 2d 3d mammography systems

Affirm® prone breast biopsy system

Selenia® dimensions® 3dimensions™ mammography system

2D 3d mammography systems

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Faradrive™ steerable sheath

Bd maxplus ™ connector minibore extension set

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Aptima hiv quant dx assay, hcv quant dx assay, and hbv quant assay software

Guedel airway

Intellicuff®

Volara system patient ciruit
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 21 Jun 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.