Horizon x-ray bone densitometer (dxa)

Official notice

21 Jun 2024

Last checked

21 Jun 2024

High riskHealth Canada
Watch Hologic

An official Health Canada notice concerns Horizon x-ray bone densitometer (dxa). During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024 Published 2024-06-21.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75719

Official notice

2024-06-21

The hazard

During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

75719

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Horizon X-Ray Bone Densitometer (Dxa), Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - HORIZON C HORIZON A HORIZON W

Risk / hazard

During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability. Recall start date: June 14, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Hologic, Horizon X-ray Bone Densitometer

Manufacturer

Not provided

Phone

Not provided

Brand: Hologic, Horizon X-ray Bone Densitometer.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type I

App estimate

high

Recall details

Recall number

75719

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Selenia dimensions 2d 3d mammography systems

Medium riskHealth Canada10 Jun 2026
CA
Recall image placeholder

Affirm® prone breast biopsy system

Medium riskHealth Canada30 Jul 2025
CA
Recall image placeholder

Selenia® dimensions® 3dimensions™ mammography system

Medium riskHealth Canada6 Sept 2024
CA
Recall image placeholder

2D 3d mammography systems

Medium riskHealth Canada7 Feb 2023
CA
Recall image placeholder

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
Recall image placeholder

Faradrive™ steerable sheath

Medium riskHealth Canada19 Jun 2026
CA
Recall image placeholder

Bd maxplus ™ connector minibore extension set

High riskHealth Canada19 Jun 2026
CA
Recall image placeholder

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Critical riskHealth Canada16 Jun 2026
CA
Recall image placeholder

Aptima hiv quant dx assay, hcv quant dx assay, and hbv quant assay software

Medium riskHealth Canada11 Jun 2026
CA
Recall image placeholder

Guedel airway

High riskHealth Canada9 Jun 2026
CA
Recall image placeholder

Intellicuff®

Medium riskHealth Canada3 Jun 2026
CA
Recall image placeholder

Volara system patient ciruit

Critical riskHealth Canada29 May 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 21 Jun 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.