No image published for this recall

No image published for this recall

Horizon® microclip™ titanium ligating clips with tape

Official notice

11 May 2023

Last checked

11 May 2023

Medium riskHealth Canada

An official Health Canada notice concerns Horizon® microclip™ titanium ligating clips with tape. Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be ligated, potentially risking injury to other distal anatomical structures. Recall start date: May 2, 2023 Published 2023-05-11.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73659

Official notice

2023-05-11

The hazard

Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be ligated, potentially risking injury to other distal anatomical structures. Recall start date: May 2, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

73659

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Horizon® Microclip™ Titanium Ligating Clips with Tape, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 005200

Risk / hazard

Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be ligated, potentially risking injury to other distal anatomical structures. Recall start date: May 2, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Teleflex Medical, Horizon® Microclip™ Titanium Ligating Clips With Tape.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

73659

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 May 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.