Histocore arcadia h

Official notice

8 May 2024

Last checked

8 May 2024

Critical riskHealth Canada

An official Health Canada notice concerns Histocore arcadia h. As part of the company's post market surveillance, they have identified an issue related to the safe usage of the HistoCore Arcadia H system. The problem is associated with the sealing of the paraffin tank. The process of paraffin solidification in the paraffin tank may lead to the deformation of the tank. This in turn, may cause the sealing of the tank to tear. Subsequent overfilling of the paraffin can cause the excess of melted paraffin to flow into the housing of the HistoCore Arcadia H. The paraffin can then soak into the insulation material. This may cause the saturated insulation material to fall onto the printed circuit board assembly and primary circuit. Because the paraffin-saturated insulation material no longer retains its flame resistance properties, this may lead to the development of toxic smoke and an internal fire in the HistoCore Arcadia H. Instruments manufactured after April 2017 do already have a reinforced tank which avoids any bending. Recall start date: May 2, 2024 Published 2024-05-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75498

Official notice

2024-05-08

The hazard

As part of the company's post market surveillance, they have identified an issue related to the safe usage of the HistoCore Arcadia H system. The problem is associated with the sealing of the paraffin tank. The process of paraffin solidification in the paraffin tank may lead to the deformation of the tank. This in turn, may cause the sealing of the tank to tear. Subsequent overfilling of the paraffin can cause the excess of melted paraffin to flow into the housing of the HistoCore Arcadia H. The paraffin can then soak into the insulation material. This may cause the saturated insulation material to fall onto the printed circuit board assembly and primary circuit. Because the paraffin-saturated insulation material no longer retains its flame resistance properties, this may lead to the development of toxic smoke and an internal fire in the HistoCore Arcadia H. Instruments manufactured after April 2017 do already have a reinforced tank which avoids any bending. Recall start date: May 2, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75498

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - HistoCore Arcadia H, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 14039354100

Risk / hazard

As part of the company's post market surveillance, they have identified an issue related to the safe usage of the HistoCore Arcadia H system. The problem is associated with the sealing of the paraffin tank. The process of paraffin solidification in the paraffin tank may lead to the deformation of the tank. This in turn, may cause the sealing of the tank to tear. Subsequent overfilling of the paraffin can cause the excess of melted paraffin to flow into the housing of the HistoCore Arcadia H. The paraffin can then soak into the insulation material. This may cause the saturated insulation material to fall onto the printed circuit board assembly and primary circuit. Because the paraffin-saturated insulation material no longer retains its flame resistance properties, this may lead to the development of toxic smoke and an internal fire in the HistoCore Arcadia H. Instruments manufactured after April 2017 do already have a reinforced tank which avoids any bending. Recall start date: May 2, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Leica Biosystems Nussloch, Histocore Arcadia H.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

critical

Recall details

Recall number

75498

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 8 May 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.