Getinge washer disinfector

Official notice

28 Aug 2025

Last checked

28 Aug 2025

Medium riskHealth Canada

An official Health Canada notice concerns Getinge washer disinfector. The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025 Published 2025-08-28.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77997

Official notice

2025-08-28

The hazard

The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77997

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Getinge Washer/Disinfector, Lot or serial number - W50037316, Model or catalogue number - 88-5

Risk / hazard

The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Getinge Disinfection, Getinge Washer Disinfector.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

77997

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 28 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.