No image published for this recall
F&p airvo 3
Official notice
29 Aug 2024
Last checked
29 Aug 2024
An official Health Canada notice concerns F&p airvo 3. This recall relates to specific versions of software of the Airvo 3, and how the device responds when flow alignment alarm 3.2.2 occurs. When this happens in Airvo 3 devices set up with high pressure oxygen (HPO) and running software version 1.5.1 or earlier, the device will deliver room air only. If this happens, a patient may experience oxygen desaturation that could lead to hypoxia. Software version 1.5.2 updates the software algorithm and ensures the target therapy continues in the event that this alarm occurs. Recall Start Date: August 8, 2024 Published 2024-08-29.
What to do now
Customer will be contacted to arrange for software update.
Recall number
75987
Official notice
2024-08-29
The hazard
The hazard
This recall relates to specific versions of software of the Airvo 3, and how the device responds when flow alignment alarm 3.2.2 occurs. When this happens in Airvo 3 devices set up with high pressure oxygen (HPO) and running software version 1.5.1 or earlier, the device will deliver room air only. If this happens, a patient may experience oxygen desaturation that could lead to hypoxia. Software version 1.5.2 updates the software algorithm and ensures the target therapy continues in the event that this alarm occurs. Recall Start Date: August 8, 2024
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Fisher And Paykel Healthcare, F&p Airvo 3.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 29 Aug 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.