Evis exera iii video system - gastrointestinal videoscope

Official notice

1 Nov 2023

Last checked

1 Nov 2023

Medium riskHealth Canada

An official Health Canada notice concerns Evis exera iii video system - gastrointestinal videoscope. The GIF-1TH190 reprocessing manual has instructions for either High Level Disinfection (HLD) reprocessing or Ethylene Oxide Gas (EOG) sterilization after endoscope cleaning. As a result of testing, Olympus discovered EOG sterilization failures in which the GIF-1TH190 failed to achieve sterilization following the GIF-1TH190's reprocessing manual. As a result of the subsequent investigation, Olympus determined that changes are required to the endoscope's channel drying steps following cleaning and prior to EOG sterilization in order to ensure effective endoscope sterilization. Accordingly, Olympus has updated the instructions for the endoscope channel's drying process and validated that the new, updated channel drying steps will result in GIF-1TH190 endoscope EOG sterilization. The revised instructions for the "dry the endoscope" in chapter 5.7 of the GIF-1TH190 reprocessing manual are provided in a supplied addendum. Recall start date: October 11, 2023 Published 2023-11-01.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74508

Official notice

2023-11-01

The hazard

The GIF-1TH190 reprocessing manual has instructions for either High Level Disinfection (HLD) reprocessing or Ethylene Oxide Gas (EOG) sterilization after endoscope cleaning. As a result of testing, Olympus discovered EOG sterilization failures in which the GIF-1TH190 failed to achieve sterilization following the GIF-1TH190's reprocessing manual. As a result of the subsequent investigation, Olympus determined that changes are required to the endoscope's channel drying steps following cleaning and prior to EOG sterilization in order to ensure effective endoscope sterilization. Accordingly, Olympus has updated the instructions for the endoscope channel's drying process and validated that the new, updated channel drying steps will result in GIF-1TH190 endoscope EOG sterilization. The revised instructions for the "dry the endoscope" in chapter 5.7 of the GIF-1TH190 reprocessing manual are provided in a supplied addendum. Recall start date: October 11, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74508

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Evis Exera III Video System - Gastrointestinal Videoscope, Lot or serial number - All lots, Model or catalog number - GIF-1TH190

Risk / hazard

The GIF-1TH190 reprocessing manual has instructions for either High Level Disinfection (HLD) reprocessing or Ethylene Oxide Gas (EOG) sterilization after endoscope cleaning. As a result of testing, Olympus discovered EOG sterilization failures in which the GIF-1TH190 failed to achieve sterilization following the GIF-1TH190's reprocessing manual. As a result of the subsequent investigation, Olympus determined that changes are required to the endoscope's channel drying steps following cleaning and prior to EOG sterilization in order to ensure effective endoscope sterilization. Accordingly, Olympus has updated the instructions for the endoscope channel's drying process and validated that the new, updated channel drying steps will result in GIF-1TH190 endoscope EOG sterilization. The revised instructions for the "dry the endoscope" in chapter 5.7 of the GIF-1TH190 reprocessing manual are provided in a supplied addendum. Recall start date: October 11, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Olympus Medical, Evis Exera Iii Video System - Gastrointestinal Videoscope.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

74508

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 1 Nov 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.