Watch Bionostics

An official Health Canada notice concerns Directcheck™ whole blood control. Instances of consistently elevated results were identified when using directCHECK™ Whole Blood Control DCGLR‑2, lot A6DLA001. A process error led to testing the lot on the wrong instrument platform, which resulted in the creation and release of incorrect Quality Control (QC) ranges on the performance range card distributed with this lot. Published 2026-05-14.

What to do now

Customer notification letter to be sent. Instructions to be provided on disposal of affected units

Recall number

82075

Official notice

2026-05-14

The hazard

Instances of consistently elevated results were identified when using directCHECK™ Whole Blood Control DCGLR‑2, lot A6DLA001. A process error led to testing the lot on the wrong instrument platform, which resulted in the creation and release of incorrect Quality Control (QC) ranges on the performance range card distributed with this lot.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Lots

A6DLA001

Skus

DCGLR‑2

Other

RA RA-82075

Model Numbers

DCGLR‑2

Recall Numbers

82075

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - A6DLA001, Model or catalogue number - DCGLR‑2

Risk / hazard

Instances of consistently elevated results were identified when using directCHECK™ Whole Blood Control DCGLR‑2, lot A6DLA001. A process error led to testing the lot on the wrong instrument platform, which resulted in the creation and release of incorrect Quality Control (QC) ranges on the performance range card distributed with this lot.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Bionostics

Manufacturer

Not provided

Phone

Not provided

Brand: Bionostics.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

82075

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Rapidlab® 348ex system

Medium riskHealth Canada2 Jul 2026
CA
Recall image placeholder

A.t.s® 4000 ts tourniquet systems single and dual hose with cpc connectors

Medium riskHealth Canada23 Jun 2026
CA
Recall image placeholder

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Medium riskHealth Canada18 Jun 2026
CA
Recall image placeholder

Cannapiece corp. Recalls wild west 96+ strawberry stampede liquid diamonds 510 vape due to a labelling issue

Critical riskHealth Canada10 Jun 2026
CA
Recall image placeholder

Neurological sponges

Medium riskHealth Canada21 May 2026
CA
Recall image placeholder

Cannapiece corp. Recalls jays - pink pennifer 0.4g all-in-one vape due to labelling issue

Critical riskHealth Canada20 May 2026
CA
Recall image placeholder

Dimension® creatinine (cre2)

Medium riskHealth Canada2 Apr 2026
CA
Recall image placeholder

Gem paks for the gem premier 5000 with iqm2

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Rapidpoint® 500 blood gas systems

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Olympus medical pk pks everest cutting forceps

Medium riskHealth Canada17 Mar 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 14 May 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.