Digital radiography mobile x-ray system dx-d 100
Official notice
12 Mar 2024
Last checked
12 Mar 2024
An official Health Canada notice concerns Digital radiography mobile x-ray system dx-d 100. A component supplier of the Agfa Dx-D 100 informed Agfa about a potential safety issue. The issue as described by the supplier is related to the steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm if the cable is cut off does not work properly. The interference occurs between the steel cable and a metal plate of the arm's carriage due to the shape of the cable's end. Recall Start Date: February 23, 2024 Published 2024-03-12.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75221
Official notice
2024-03-12
The hazard
The hazard
A component supplier of the Agfa Dx-D 100 informed Agfa about a potential safety issue. The issue as described by the supplier is related to the steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm if the cable is cut off does not work properly. The interference occurs between the steel cable and a metal plate of the arm's carriage due to the shape of the cable's end. Recall Start Date: February 23, 2024
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Agfa N.v, Digital Radiography Mobile X-ray System Dx-d 100.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 12 Mar 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.