No image published for this recall
Defigard touch7 monitor and defibrillator
Official notice
23 Nov 2023
Last checked
23 Nov 2023
An official Health Canada notice concerns Defigard touch7 monitor and defibrillator. In rare cases, an ECG failure affecting impedance measurement via defibrillation electrodes has led to the Defigard Touch-7 falsely reports an electrode fault, resulting in inoperative defibrillation. Recall start date: November 14, 2023 Published 2023-11-23.
What to do now
Contact the manufacturer if you require additional information.
Recall number
74632
Official notice
2023-11-23
The hazard
The hazard
In rare cases, an ECG failure affecting impedance measurement via defibrillation electrodes has led to the Defigard Touch-7 falsely reports an electrode fault, resulting in inoperative defibrillation. Recall start date: November 14, 2023
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Schiller Medical, Defigard Touch7 Monitor And Defibrillator.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Defigard touch 7

Defigard touch7

Defigard touch7

Atlan a350
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 23 Nov 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.