No image published for this recall

No image published for this recall

Dako coverstainers

Official notice

11 Apr 2024

Last checked

11 Apr 2024

Medium riskHealth Canada

An official Health Canada notice concerns Dako coverstainers. Lower front door of coverstainer instrument damaged due to hinges not being able to support weight of the door assembly. This leads to cracks in the plexiglass of the lower front door and causes pieces to break off in the top left/right corners of the assembly. Recall start date: April 2, 2024 Published 2024-04-11.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75383

Official notice

2024-04-11

The hazard

Lower front door of coverstainer instrument damaged due to hinges not being able to support weight of the door assembly. This leads to cracks in the plexiglass of the lower front door and causes pieces to break off in the top left/right corners of the assembly. Recall start date: April 2, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75383

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Coverstainer, Dako Coverstainer Cpo, Lot or serial number - CS10030,CS10040,CS10026, Model or catalogue number - CS100

Product 2

Affected products - Dako Coverstainer, Dako Coverstainer Cpo, Lot or serial number - CS10030,CS10040,CS10026, Model or catalogue number - CS100

Risk / hazard

Lower front door of coverstainer instrument damaged due to hinges not being able to support weight of the door assembly. This leads to cracks in the plexiglass of the lower front door and causes pieces to break off in the top left/right corners of the assembly. Recall start date: April 2, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Gap, Dako Coverstainers.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75383

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 Apr 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.