No image published for this recall

No image published for this recall

Ct scan and high capacity table

Official notice

15 Mar 2023

Last checked

15 Mar 2023

Medium riskHealth Canada

An official Health Canada notice concerns Ct scan and high capacity table. During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop. Recall start date: June 6, 2022 Published 2023-03-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64292

Official notice

2023-03-15

The hazard

During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop. Recall start date: June 6, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64292

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - High Capacity Table, Lot or serial number - 00000222467HM3 00000257352HM5 00000270176HM1 00000256160HM3 00000293344HM8 359928HM9 CL3GT2100005HM CL3GT1800006HM, Model or catalogue number - 5272966 5272966-2 5272966-3

Product 2

Affected products - Gt2000x/Vt2000x, Lot or serial number - 00000413151HM2 CL1GT1600075HM CL1GT1600061HM 00000429212HM4 00000413848HM3 CL1GT1600038HM 000394049HM1 00000429230HM6 00000341402HM6 CL1GT1900057HM CL1GT2100036HM CL1GT1800039HM CL1GT1900060HM CL1GT1900040HM CL1GT1900042HM CL1GT1800049HM CL1GT2100015HM HDLVX1900005CN CL1GT2100091HM CL1GT1900058HM CL1GT1800077HM CL1GT2000019HM CL1GT1900035HM CL1GT2100002HM CL1GT2000062HM CL1GT1900082HM 00000386645HM6 CL1GT1600029HM 413150HM4 CL1GT2100019HM, Model or catalogue number - 5380966

Risk / hazard

During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop. Recall start date: June 6, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: GE Hangwei Medical, Ct Scan And High Capacity Table.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64292

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Image: Health Canada recall notice

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Cyberknife® treatment delivery system

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Unauthorized soft-shelled hyperbaric chambers may pose serious health risks 2026-06-16

Critical riskHealth Canada16 Jun 2026
CA
No image published for this recall

Harmony™ delivery catheter system

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Impella controller

Critical riskHealth Canada13 May 2026
CA
No image published for this recall

Atlan a350

High riskHealth Canada4 May 2026
CA
No image published for this recall

Automated impella® controller

Critical riskHealth Canada28 Apr 2026
CA
No image published for this recall

Cook medical beacon® tip centimeter sizing catheter

Medium riskHealth Canada28 Apr 2026
CA
No image published for this recall

Cardinal health™ angiography pack and catheterization packs

High riskHealth Canada15 Apr 2026
CA
No image published for this recall

Soft-vu and accu-vu 4 fr angiographic catheters

Medium riskHealth Canada18 Mar 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 15 Mar 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.