No image published for this recall

No image published for this recall

Codman® disposable perforator

Official notice

22 Oct 2025

Last checked

22 Oct 2025

Medium riskHealth Canada

An official Health Canada notice concerns Codman® disposable perforator. On April 2025, Integra issued a voluntary recall for the 14mm Codman® Disposable Perforators due to an inadequate weld (proud weld) on specific lots of 14mm Codman® Disposable Perforators that can potentially cause the product to disassemble (break/separate). To date, there have been no complaints or adverse events reported in relation to the 9mm and 11mm Codman® Disposable Perforators disassembly. Out of an abundance of caution, as the 9 and 11 mm perforators are manufactured using the same equipment and operating parameters that affected the 14mm Perforator, Integra has made the decision to expand the scope of the field action to include the 9mm and 11mm perforators. Published 2025-10-22.

What to do now

Contact the manufacturer if you require additional information

Recall number

77368

Official notice

2025-10-22

The hazard

On April 2025, Integra issued a voluntary recall for the 14mm Codman® Disposable Perforators due to an inadequate weld (proud weld) on specific lots of 14mm Codman® Disposable Perforators that can potentially cause the product to disassemble (break/separate). To date, there have been no complaints or adverse events reported in relation to the 9mm and 11mm Codman® Disposable Perforators disassembly. Out of an abundance of caution, as the 9 and 11 mm perforators are manufactured using the same equipment and operating parameters that affected the 14mm Perforator, Integra has made the decision to expand the scope of the field action to include the 9mm and 11mm perforators.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77368

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Codman® Disposable Perforator, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 26-1221

Product 2

Affected products - Codman® Disposable Perforator, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 26-1222

Product 3

Affected products - Codman® Disposable Perforator, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 26-1223

Risk / hazard

On April 2025, Integra issued a voluntary recall for the 14mm Codman® Disposable Perforators due to an inadequate weld (proud weld) on specific lots of 14mm Codman® Disposable Perforators that can potentially cause the product to disassemble (break/separate). To date, there have been no complaints or adverse events reported in relation to the 9mm and 11mm Codman® Disposable Perforators disassembly. Out of an abundance of caution, as the 9 and 11 mm perforators are manufactured using the same equipment and operating parameters that affected the 14mm Perforator, Integra has made the decision to expand the scope of the field action to include the 9mm and 11mm perforators.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Integra Lifesciences Production, Codman® Disposable Perforator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

77368

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Codman microsensor® basic kits and cerelink™ icp sensor basic kits

Medium riskHealth Canada30 Dec 2025
CA
No image published for this recall

Codman® surgical strips and patties x-ray detectable

Medium riskHealth Canada24 Oct 2024
CA
No image published for this recall

Codman bactiseal evd catheter set

Critical riskHealth Canada29 May 2024
CA
No image published for this recall

Cerelink™ icp extension cable

Medium riskHealth Canada24 Aug 2023
CA
No image published for this recall

Cerelink icp monitor

Medium riskHealth Canada18 Jul 2022
CA
No image published for this recall

Hotline IV fluid warmer system & accessories

High riskHealth Canada6 Jul 2026
CA
No image published for this recall

Hybrid hydrophilic microguidewire

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Intramedullary bone saw blade assembly

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Glidesheath slender® introducer sheath

Medium riskHealth Canada2 Jul 2026
CA
No image published for this recall

Infinion™ cx contact lead kit

Critical riskHealth Canada24 Jun 2026
CA
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Image: Health Canada recall notice

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
No image published for this recall

Halyard* purezero* ultraviolet* nitrile exam gloves

Medium riskHealth Canada17 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 22 Oct 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.