No image published for this recall

No image published for this recall

Broselow™ hinkle pediatric emergency system sets

Official notice

19 Dec 2025

Last checked

19 Dec 2025

Critical riskHealth Canada

An official Health Canada notice concerns Broselow™ hinkle pediatric emergency system sets. AirLife™ has identified three medication-related errors in the content printed on the affected product version of the AirLife™ Broselow™ Rainbow Tape. Incorrect dosage of ketamine, flumazenil, and vecuronium is listed on the color zones of the Broselow™ tape which could potentially lead to an adverse event. Published 2025-12-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81387

Official notice

2025-12-19

The hazard

AirLife™ has identified three medication-related errors in the content printed on the affected product version of the AirLife™ Broselow™ Rainbow Tape. Incorrect dosage of ketamine, flumazenil, and vecuronium is listed on the color zones of the Broselow™ tape which could potentially lead to an adverse event.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

0004338439, 0004329078, 0004329079, 0004325583, 0004331255, 0004329077, 0004333368, 0004335537, 0004317699, 0004333593, 0004333171

Skus

7700RE, 7730IALS

Other

RA RA-81387

Model Numbers

7700RE, 7730IALS

Recall Numbers

81387

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004338439, Model or catalogue number - 7700RE

Product 2

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004329078, Model or catalogue number - 7700RE

Product 3

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004329079, Model or catalogue number - 7700RE

Product 4

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004325583, Model or catalogue number - 7700RE

Product 5

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004331255, Model or catalogue number - 7700RE

Product 6

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004329077, Model or catalogue number - 7700RE

Product 7

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004333368, Model or catalogue number - 7700RE

Product 8

Affected products - Broselow™/Hinkle Pediatric Emergency System Emergency Tape, Lot or serial number - 0004335537, Model or catalogue number - 7700RE

Product 9

Affected products - Broselow™/Hinkle Pediatric Emergency System Kit, Lot or serial number - 0004317699, Model or catalogue number - 7730IALS

Product 10

Affected products - Broselow™/Hinkle Pediatric Emergency System Kit, Lot or serial number - 0004333593, Model or catalogue number - 7730IALS

Product 11

Affected products - Broselow™/Hinkle Pediatric Emergency System Kit, Lot or serial number - 0004333171, Model or catalogue number - 7730IALS

Product 12

Lot or serial number - 0004338439, Model or catalogue number - 7700RE

Product 13

Lot or serial number - 0004329078, Model or catalogue number - 7700RE

Product 14

Lot or serial number - 0004329079, Model or catalogue number - 7700RE

Product 15

Lot or serial number - 0004325583, Model or catalogue number - 7700RE

Product 16

Lot or serial number - 0004331255, Model or catalogue number - 7700RE

Product 17

Lot or serial number - 0004329077, Model or catalogue number - 7700RE

Product 18

Lot or serial number - 0004333368, Model or catalogue number - 7700RE

Product 19

Lot or serial number - 0004335537, Model or catalogue number - 7700RE

Product 20

Lot or serial number - 0004317699, Model or catalogue number - 7730IALS

Product 21

Lot or serial number - 0004333593, Model or catalogue number - 7730IALS

Product 22

Lot or serial number - 0004333171, Model or catalogue number - 7730IALS

Risk / hazard

AirLife™ has identified three medication-related errors in the content printed on the affected product version of the AirLife™ Broselow™ Rainbow Tape. Incorrect dosage of ketamine, flumazenil, and vecuronium is listed on the color zones of the Broselow™ tape which could potentially lead to an adverse event.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Airlife.

Source & status

Source

Health Canada

Tier

beta

Category

HEALTH

Classification

Type I

App estimate

critical

Recall details

Recall number

81387

Classification

Type I

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 19 Dec 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.